Complete PTR Quadrupole MS

Photo of system

Photo
Photograph showing the PTR section with its multiple plate electrodes and the electron impact source on the right side.

Photo
Close-up photograph of the glow discharge source and PTR section, with cover removed.

At the client's request, we have designed and delivered a research quadrupole-based PTR-MS (proton transfer reactor mass spectrometer).

The instrument is now at Horiba in Japan, where its main function will be to explore the detection of low level components in automotive exhausts in real time during engine monitoring.

The instrument has been designed with an 'open architecture' that permits a researcher to change the configuration of the PTR front end to accommodate various experiments.

PTR Method Using H3O+ (Ionisation of analyte by proton transfer from H3O+)

In this mode a primary beam, H3O+ (or 'Hydronium'), is created within the hollow cathode glow discharge source. The beam is passed into the PTR section where analyte is introduced. General components in air are not ionised by the hydronium beam, but some inorganic gases and most organic gases are ionised in a 'soft' manner that preserves the molecule.

The technique requires that the primary beam species are always in excess compared to the analyte, which is dilute. Up to (approximately) 50ppm of an analyte can be analysed without the hydronium population being reduced significantly (thus ensuring a linear response).

For more information on how this method works, please refer to: Proton Transfer Reactions Using H3O+ Ion Beams

Example Data: H2S detection in PTR Mode

In a conventional mass spectrometer, detecting S in the presence of O2 is not trivial. The main isotope of S is 32S with an exact mass of 31.972. The most abundant isotope of oxygen is 16O, and the exact mass of 16O2 is 31.989. Thus, both ions appear at mass 32. This is known as an 'isobaric mass interference'.

The H2S ion occurs at mass 34 (1H232 S). Unfortunately, one of the (isotopic) ions of O2 is 16 O18O+ , and this also has a mass of 34. The 'exact' masses of these two molecules are 33.9877 and 33.994 respectively, and thus a mass spectrometer would need a resolving power of 5,400 to separate the two mass peaks. This is certainly possible (though not with a quadrupole MS), but it has implications for the size and cost of such an instrument.

The PTR-TOF instrument offers an alternative way to overcome such 'mass interference'.

In the PTR-TOF (proton transfer reaction TOF), it is possible to employ an ionisation scheme that uses the ion H3O+ to ionise the compounds of interest, by a method known as "proton transfer". H2O has a 'proton affinity' of 691 kJ/mol. This is less than H2S, which has a proton affinity of 709 kJ/mol, and so in a collision H2S will remove the proton from H3O+ and become ionised. By contrast, O2 has a proton affinity of 421 kJ/mol, and will not be ionised by H3O+. Thus it should be possible to distinguish completely between H 2S and O2, because only the H2S will be ionised in this particular PTR scheme.


Graph

This is mass spectral data was acquired from a sample of H2S (50ppm) in Nitrogen (log scale Y-axis)

600,000 c/s of were recorded at mass 35 (H2S+), i.e. 12,000 c/s per ppm. This is equivalent to a detection sensitivity of 50ppb in 0.1 second.


Graph

This is data acquired immediately after the H2S sample, from room air.

The signal at mass 35 has disappeared.


Example Data: Methanol Detection in PTR Mode

Graph

Methanol (CH3OH) has a proton affinity of 754kJ/mol compared to 691kJ/mol for water. Thus, methanol can be ionised by proton transfer.

The adjacent data was acquired from a gas mixture of 10ppm methanol in nitrogen. The molecular mass of methanol is 32, but with an added proton the appearance mass is at 33. Greater than 9 million c/s second were recorded at mass 33.


Chemical Ionisation by Ion Molecule Reaction

(Ionisation of analyte by ion - molecule reactions)

In this mode, the glow discharge is run on an inert gas such as argon, xenon or krypton, instead of N2/H2O mixture. The primary ion produced from the glow discharge source has a specific energy in its ionised state. If it impacts a molecule in the collision cell and that molecule requires less energy to be ionised, then that molecule can become ionised and the primary ion will convert to a neutral particle. The sequence is as follows:

Kr+ (13.99 eV) + CO → CO+ (13.98 eV) + Kr
(Chemical Ionisation by Ion-Molecule Reaction)

However, if the molecule requires more energy than is available in the rare gas ion, the molecule remains neutral, e.g.

Kr+ (13.99 eV) + N2 → Kr+ + N2
(No Chemical Ionisation, because 15.5 eV is required to ionise N2).

For further details refer to Chemical Ionisation by Ion-Molecule Reaction

Example Data in Ion-Molecule Mode (Ar+ Ion Beam)

Argon has an ionisation energy of 15.76eV. SO2 has an ionisation energy of 12.32eV, thus should be ionised in a collision with Ar+.

Graph

Adjacent is data acquired from a 50ppm SO2 in Nitrogen sample. 100,000 c/s were at mass 65 for this analyte using Ar+


Electron Impact Ionisation (EI Mode)

The instrument features a conventional electron impact ionisation source in the high vacuum section. When an analyte of interest cannot be ionised by the proton transfer method, the glow discharge source may be switched off and the electron impact source activated.

Electron impact ionisation is a very common method, and a large database of EI spectra exists (NIST database). All components are ionised by this method, but the excess of energy does result in fragmentation. At times this can be useful for identifying a compound.

Example Data in EI-Mode

Graph

The adjacent spectrum was acquired from a calibration gas mixture containing 2.6% Hydrogen. The signal at mass 2 was 3.6 million counts per second in EI mode, and this equates to a detection sensitivity of 0.005% (limited by background count at mass 2 due to hydrogen in the system).


Top